Assessing the performance of a robust multiparametric wearable patch integrating silicon-based sensors for real-time continuous monitoring of sweat biomarkers.

Détails

Ressource 1Télécharger: 39018979.pdf (4587.46 [Ko])
Etat: Public
Version: Final published version
Licence: CC BY 4.0
ID Serval
serval:BIB_271D2D9BA3C0
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Assessing the performance of a robust multiparametric wearable patch integrating silicon-based sensors for real-time continuous monitoring of sweat biomarkers.
Périodique
Biosensors & bioelectronics
Auteur⸱e⸱s
Rovira M., Lafaye C., Demuru S., Kunnel B.P., Aymerich J., Cuenca J., Serra-Graells F., Margarit-Taulé J.M., Haque R., Saubade M., Fernández-Sánchez C., Jimenez-Jorquera C.
ISSN
1873-4235 (Electronic)
ISSN-L
0956-5663
Statut éditorial
Publié
Date de publication
15/10/2024
Peer-reviewed
Oui
Volume
262
Pages
116560
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
The development of wearable devices for sweat analysis has experienced significant growth in the last two decades, being the main focus the monitoring of athletes health during workouts. One of the main challenges of these approaches has been to attain the continuous monitoring of sweat for time periods over 1 h. This is the main challenge addressed in this work by designing an analytical platform that combines the high performance of potentiometric sensors and a fluidic structure made of a plastic fabric into a multiplexed wearable device. The platform comprises Ion-Sensitive Field-Effect Transistors (ISFETs) manufactured on silicon, a tailor-made solid-state reference electrode, and a temperature sensor integrated into a patch-like polymeric substrate, together with the component that easily collects and drives samples under continuous capillary flow to the sensor areas. ISFET sensors for measuring pH, sodium, and potassium ions were fully characterized in artificial sweat solutions, providing reproducible and stable responses. Then, the real-time and continuous monitoring of the biomarkers in sweat with the wearable platform was assessed by comparing the ISFETs responses recorded during an 85-min continuous exercise session with the concentration values measured using commercial Ion-Selective Electrodes (ISEs) in samples collected at certain times during the session. The developed sensing platform enables the continuous monitoring of biomarkers and facilitates the study of the effects of various real working conditions, such as cycling power and skin temperature, on the target biomarker concentration levels.
Mots-clé
Wearable Electronic Devices, Sweat/chemistry, Biosensing Techniques/instrumentation, Humans, Silicon/chemistry, Biomarkers/analysis, Transistors, Electronic, Equipment Design, Sodium/analysis, Potassium/analysis, Hydrogen-Ion Concentration, Monitoring, Physiologic/instrumentation, Capillary-based fluidics, Fitness management, Healthcare, Microfabricated sensor, Sweat biomarker, Wearable sensor
Pubmed
Web of science
Open Access
Oui
Création de la notice
19/07/2024 10:25
Dernière modification de la notice
09/08/2024 14:53
Données d'usage