Relation between intraoperative EMG values and final pedicle screw position as seen on CT images

Détails

ID Serval
serval:BIB_24B9084CD82E
Type
Actes de conférence (partie): contribution originale à la littérature scientifique, publiée à l'occasion de conférences scientifiques, dans un ouvrage de compte-rendu (proceedings), ou dans l'édition spéciale d'un journal reconnu (conference proceedings).
Sous-type
Abstract (résumé de présentation): article court qui reprend les éléments essentiels présentés à l'occasion d'une conférence scientifique dans un poster ou lors d'une intervention orale.
Collection
Publications
Institution
Titre
Relation between intraoperative EMG values and final pedicle screw position as seen on CT images
Titre de la conférence
Annual meeting of the Swiss Society of Orthopaedics and Traumatology
Auteur⸱e⸱s
Kulik G., McManus J. R., Pralong E., Debatisse D., Schizas C.
Adresse
St. Gallen - Switzerland, June 30 - July 2, 2010
ISBN
1424-7860
Statut éditorial
Publié
Date de publication
2010
Peer-reviewed
Oui
Volume
140
Série
Swiss Medical Weekly
Pages
29S
Langue
anglais
Notes
Meeting Abstract
Résumé
Introduction: Intraoperative EMG based neurophysiological monitoring is increasingly used to assist pedicle screw insertion. We carried out a study comparing the final screw position in the pedicle measured on CT images in relation to its corresponding intraoperative muscle compound action potential (CMAP) values.
Material and methods: A total of 189 screws were inserted in thoracolumbar spines of 31 patients during instrumented fusion under EMG control. An observer, blinded to the CMAP value, assessed the horizontal and vertical 'screw edge to pedicle edge' distance perpendicular to the longitudinal axis of the screw on reformatted CT reconstructions using OsiriX software. These distances were analysed with their corresponding CMAP values. Data from 62 thoracic and 127 lumbar screws were processed separately. Interobserver reliability of distance measurements was assessed.
Results: No patient suffered neurological injury secondary to screw insertion. Distance measurements were reliable (paired t-test, P = 0.13/0.98 horizontal/vertical). Two screws had their position altered due to low CMAP values suggesting close proximity of nerve tissue. Seventy five percent of screws had CMAP results above 10mA and had an average distance of 0.35cm (SD 0.23) horizontally and 0.46cm (SD 0.26) vertically from the pedicle edge. Additional 12% had a distance from the edge of the pedicle less than 0mm indicating cortical breach but had CMAP values above 10mA. A poor correlation between CMAP values and screw position was found.
Discussion: In this study CMAP values above 10mA indicated correct screw position in the majority of cases. The zone of 10-20mA CMAP carries highest risk of a misplaced screw despite high CMAP value (17% of screws this CMAP range). In order to improve accuracy of EMG predictive value further research is warranted including improvement of probing techniques.
Web of science
Création de la notice
14/10/2010 12:58
Dernière modification de la notice
20/08/2019 14:03