Principes de la mesure de la depense energetique. [Principles of the measurement of energy expenditure]

Détails

ID Serval
serval:BIB_2198620EA349
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Principes de la mesure de la depense energetique. [Principles of the measurement of energy expenditure]
Périodique
Agressologie
Auteur⸱e⸱s
Thorin  D., Chiolero  R.
ISSN
0002-1148 (Print)
Statut éditorial
Publié
Date de publication
01/1990
Volume
31
Numéro
1
Pages
59-61
Notes
English Abstract
Journal Article --- Old month value: Jan
Résumé
Numerous methods can be used to assess energy expenditure (EE). The Fick method allows an easy calculation of EE in critically ill patients with Swan-Ganz catheters. Indirect calorimetry can be performed using either closed circuit or open circuit techniques. The latter is accurate when the patient is breathing air or low-FiO2 gaz mixture, but inaccurate at high FiO2. The accuracy of instruments, the presence of a leak-free system and the mathematical manipulations are critical points. Calorimetric studies performed in patients receiving high FiO2 must be interpreted with caution.
Mots-clé
Calorimetry/methods *Energy Metabolism Female Humans Male Oxygen Consumption Reference Values