Anatomic shoulder arthroplasty: consequences of humeral head malposition

Détails

ID Serval
serval:BIB_1BB0EA5DDF70
Type
Actes de conférence (partie): contribution originale à la littérature scientifique, publiée à l'occasion de conférences scientifiques, dans un ouvrage de compte-rendu (proceedings), ou dans l'édition spéciale d'un journal reconnu (conference proceedings).
Sous-type
Abstract (résumé de présentation): article court qui reprend les éléments essentiels présentés à l'occasion d'une conférence scientifique dans un poster ou lors d'une intervention orale.
Collection
Publications
Institution
Titre
Anatomic shoulder arthroplasty: consequences of humeral head malposition
Titre de la conférence
22th International Congress of Shoulder and Elbow Surgery
Auteur⸱e⸱s
Farron A., Terrier A.
Adresse
Madrid, Espagne, 16-19 September 2009
Statut éditorial
Publié
Date de publication
2009
Série
Book of Abstracts
Pages
85
Langue
anglais
Résumé
Purpose
Third generation anatomic total shoulder prostheses offer a wide range of adaptability (size, thickness, retroversion and offset of the humeral head, cervico-diaphyseal angle) in order to reproduce anatomy and biomechanics of the shoulder as normal as possible. The large variability of the implants may also induce malposition. Our goal was to analyse the consequences of a humeral head malposition, which is one of the most frequent placement errors.
Material and Methods
A 3D finite element model of the glenohumeral joint, including the rotator cuff muscles and the deltoid, was used with the Aequalis anatomic prosthesis. Active abduction was simulated. Three humeral head placements were compared : anatomic positioning (A), 5 mm inferior positioning (B), 5 mm superior positioning (C). The effect of humeral head malposition was evaluated through the following quantities : the range of motion free of impingements, the glenohumeral contact pattern, and the stress within the polyethylene and the cement.
Results
Inferior positioning (B) of the humeral head produced a superior impingement before 90° of abduction, an inferior eccentric contact point on the glenoid, and 165% increase of cement stress. Superior positioning (C) of the humeral head produced a postero-superior eccentric contact point on the glenoid, 300% increase of glenohumeral contact pressure, 450% increase of polyethylene stress, and 207% increase of cement stress.
Conclusion
Malposition of the humeral head of anatomic prostheses induces biomechanical consequences that may preclude the glenoid survival. Particular attention must be paid to reproduce the humeral anatomy as normal as possible.