Metallic renal artery MR imaging stent: artifact-free lumen visualization with projection and standard renal MR angiography.

Détails

ID Serval
serval:BIB_0DEC8889B9A3
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Metallic renal artery MR imaging stent: artifact-free lumen visualization with projection and standard renal MR angiography.
Périodique
Radiology
Auteur⸱e⸱s
Spuentrup E., Ruebben A., Stuber M., Günther R.W., Buecker A.
ISSN
0033-8419[print], 0033-8419[linking]
Statut éditorial
Publié
Date de publication
2003
Volume
227
Numéro
3
Pages
897-902
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
A cardiac-triggered free-breathing three-dimensional (3D) balanced fast field-echo projection renal magnetic resonance (MR) angiographic sequence was investigated for in-stent lumen visualization of a dedicated metallic renal artery stent. Fourteen prototype stents were deployed in the renal arteries of six pigs (in two pigs, three stents were deployed). Projection renal MR angiography was compared with standard contrast material-enhanced 3D breath-hold MR angiography. Artifact-free in-stent lumen visualization was achieved with both projection MR angiography and contrast-enhanced MR angiography. These promising results warrant further studies for visualization of in-stent restenosis.
Mots-clé
Alloys, Animals, Artifacts, Contrast Media, Equipment Design, Imaging, Three-Dimensional, Magnetic Resonance Angiography/methods, Renal Artery/radiography, Renal Artery Obstruction/radiography, Renal Artery Obstruction/therapy, Stents, Swine