Accuracy of a new transmittance-reflectance pulse oximetry sensor in critically ill neonates.

Détails

ID Serval
serval:BIB_0B81874CBD64
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Accuracy of a new transmittance-reflectance pulse oximetry sensor in critically ill neonates.
Périodique
Critical care medicine
Auteur⸱e⸱s
Fanconi S., Tschupp A.
ISSN
0090-3493
Statut éditorial
Publié
Date de publication
1994
Peer-reviewed
Oui
Volume
22
Numéro
7
Pages
1142-6
Langue
anglais
Notes
Publication types: Comparative Study ; Journal Article - Publication Status: ppublish
Résumé
OBJECTIVE: To test the accuracy of a new pulse oximeter sensor based on transmittance and reflectance. This sensor makes transillumination of tissue unnecessary and allows measurements on the hand, forearm, foot, and lower limb. DESIGN: Prospective, open, nonrandomized criterion standard study. SETTING: Neonatal intensive care unit, tertiary care center. PATIENTS: Sequential sample of 54 critically ill neonates (gestational age 27 to 42 wks; postnatal age 1 to 28 days) with arterial catheters in place. MEASUREMENTS AND MAIN RESULTS: A total of 99 comparisons between pulse oximetry and arterial saturation were obtained. Comparison of femoral or umbilical arterial blood with transcutaneous measurements on the lower limb (n = 66) demonstrated an excellent correlation (r2 = .96). The mean difference was +1.44% +/- 3.51 (SD) % (range -11% to +8%). Comparison of the transcutaneous values with the radial artery saturation from the corresponding upper limb (n = 33) revealed a correlation coefficient of 0.94 with a mean error of +0.66% +/- 3.34% (range -6% to +7%). The mean difference between noninvasive and invasive measurements was least with the test sensor on the hand, intermediate on the calf and arm, and greatest on the foot. The mean error and its standard deviation were slightly larger for arterial saturation values < 90% than for values > or = 90%. CONCLUSION: Accurate pulse oximetry saturation can be acquired from the hand, forearm, foot, and calf of critically ill newborns using this new sensor.
Mots-clé
Arm, Confidence Intervals, Critical Illness, Evaluation Studies as Topic, Humans, Infant, Newborn, Leg, Linear Models, Oximetry, Oxygen, Transillumination