Coronary magnetic resonance angiography for assessment of the stent lumen: a phantom study.

Détails

ID Serval
serval:BIB_09D0CC4D320A
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Coronary magnetic resonance angiography for assessment of the stent lumen: a phantom study.
Périodique
Journal of Cardiovascular Magnetic Resonance
Auteur⸱e⸱s
Maintz D., Botnar R.M., Fischbach R., Heindel W., Manning W.J., Stuber M.
ISSN
1097-6647[print], 1097-6647[linking]
Statut éditorial
Publié
Date de publication
2002
Volume
4
Numéro
3
Pages
359-367
Langue
anglais
Notes
Publication types: Comparative Study ; Evaluation Studies ; Journal Article ; Research Support, Non-U.S. Gov't
Publication Status: ppublish
Résumé
PURPOSE: To evaluate the feasibility of visualizing the stent lumen using coronary magnetic resonance angiography in vitro. MATERIAL AND METHODS: Nineteen different coronary stents were implanted in plastic tubes with an inner diameter of 3 mm. The tubes were positioned in a plastic container filled with gel and included in a closed flow circuit (constant flow 18 cm/sec). The magnetic resonance images were obtained with a dual inversion fast spin-echo sequence. For intraluminal stent imaging, subtraction images were calculated from scans with and without flow. Subsequently, intraluminal signal properties were objectively assessed and compared. RESULTS: As a function of the stent type, various degrees of in-stent signal attenuation were observed. Tantalum stents demonstrated minimal intraluminal signal attenuation. For nitinol stents, the stent lumen could be identified, but the intraluminal signal was markedly reduced. Steel stents resulted in the most pronounced intraluminal signal voids. CONCLUSIONS: With the present technique, radiofrequency penetration into the stents is strongly influenced by the stent material. Thesefindings may have important implicationsforfuture stent design and stent imaging strategies.
Mots-clé
Blood Vessel Prosthesis Implantation, Coronary Angiography/instrumentation, Coronary Artery Disease/diagnosis, Coronary Artery Disease/surgery, Coronary Vessels/surgery, Equipment Design/instrumentation, Feasibility Studies, Humans, Magnetic Resonance Angiography/instrumentation, Phantoms, Imaging, Signal Processing, Computer-Assisted/instrumentation, Stents
Pubmed
Web of science
Création de la notice
02/03/2010 17:04
Dernière modification de la notice
20/08/2019 13:31
Données d'usage