Errors and limitations of the multimodality checking methods of defective spinal intrathecal pump systems. Case report.

Détails

ID Serval
serval:BIB_07427C845DED
Type
Article: article d'un périodique ou d'un magazine.
Sous-type
Etude de cas (case report): rapporte une observation et la commente brièvement.
Collection
Publications
Titre
Errors and limitations of the multimodality checking methods of defective spinal intrathecal pump systems. Case report.
Périodique
Paraplegia
Auteur⸱e⸱s
Schurch B.
ISSN
0031-1758 (Print)
ISSN-L
0031-1758
Statut éditorial
Publié
Date de publication
1993
Volume
31
Numéro
9
Pages
611-615
Langue
anglais
Notes
Publication types: Case Reports ; Journal ArticlePublication Status: ppublish
Résumé
The loss of benefit from intrathecal baclofen, with increased spasticity and a discrepancy between the residual and the calculated volume content (underinfusion), made us suspect dysfunction of the intrathecal baclofen infusion in a paraplegic patient. Although all possible usual checking methods were used, no failure in the drug administration device (DAD) could be found. Despite some benefit from increasing the daily dose of baclofen, it has not been possible to control the spastic symptoms. Surgery was therefore decided upon, and a small leak at the end of the catheter tip was discovered. Surgeons and physicians should be aware that checking methods of DAD cannot exclude failure of the system. Their errors and limitations are discussed.
Mots-clé
Adult, Baclofen/administration & dosage, Baclofen/therapeutic use, Equipment Failure, Humans, Infusion Pumps, Implantable, Injections, Spinal/instrumentation, Male, Muscle Spasticity/drug therapy, Muscle Spasticity/etiology, Spinal Cord Injuries/complications, Spinal Cord Injuries/drug therapy
Pubmed
Web of science