Optimized venous return with a self-expanding cannula: from computational fluid dynamics to clinical application.

Détails

ID Serval
serval:BIB_055E50D28AAE
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Optimized venous return with a self-expanding cannula: from computational fluid dynamics to clinical application.
Périodique
Interactive cardiovascular and thoracic surgery
Auteur⸱e⸱s
Mueller X.M., Mallabiabrena I., Mucciolo G., von Segesser L.K.
ISSN
1569-9285[electronic]
Statut éditorial
Publié
Date de publication
2002
Volume
1
Numéro
1
Pages
23-7
Langue
anglais
Notes
Publication types: Journal Article - Publication Status: ppublish
Résumé
The Smart canula concept allows for collapsed cannula insertion, and self-expansion within a vein of the body. (A) Computational fluid dynamics, and (B) bovine experiments (76+/-3.8 kg) were performed for comparative analyses, prior to (C) the first clinical application. For an 18F access, a given flow of 4 l/min (A) resulted in a pressure drop of 49 mmHg for smart cannula versus 140 mmHg for control. The corresponding Reynolds numbers are 680 versus 1170, respectively. (B) For an access of 28F, the maximal flow for smart cannula was 5.8+/-0.5 l/min versus 4.0+/-0.1 l/min for standard (P<0.0001), for 24F 5.5+/-0.6 l/min versus 3.2+/-0.4 l/min (P<0.0001), and for 20F 4.1+/-0.3 l/min versus 1.6+/-0.3 l/min (P<0.0001). The flow obtained with the smart cannula was 270+/-45% (20F), 172+/-26% (24F), and 134+/-13% (28F) of standard (one-way ANOVA, P=0.014). (C) First clinical application (1.42 m2) with a smart cannula showed 3.55 l/min (100% predicted) without additional fluids. All three assessment steps confirm the superior performance of the smart cannula design.
Pubmed
Création de la notice
14/02/2008 15:15
Dernière modification de la notice
20/08/2019 13:27
Données d'usage