Temporary lung support using an intravascular gas exchanger
Détails
ID Serval
serval:BIB_039171155F08
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Temporary lung support using an intravascular gas exchanger
Périodique
Thoracic and Cardiovascular Surgeon
ISSN
0171-6425
Statut éditorial
Publié
Date de publication
06/1992
Peer-reviewed
Oui
Volume
40
Numéro
3
Pages
121-5
Notes
Journal Article --- Old month value: Jun
Résumé
Severe respiratory failure may be difficult to overcome by conventional mechanical ventilation. As an alternative to the very demanding lung support using various modalities of extracorporeal circulation (ECMO, ECLA, etc.) we evaluated an intravascular gas exchanger (IVOX) in a "reversible" bovine model. Several degrees of hypoventilation were studied with and without intravascular gas exchange in 5 endotracheal intubated, anaesthetized, and volume-controlled ventilated animals (body weight 73 +/- 4 kg). After systematical heparinization (300 IU/kg body weight) the animals were instrumented with EKG electrodes, thermodilution pulmonary artery catheter with continuous measurement of O2-saturation, central venous and femoral arterial catheter, etc. The intravascular gas exchanger made from siliconized microporous polypropylene hollow fibres was placed in the caval veins under radiofluoroscopic control. The following scenarios were studied without and with intravascular gas exchange (gas inlet 100% O2): Normoventilation (with 14-20 strokes/min) at F10(2) 0.50 and at F10(2) 0.21; Hypoventilation (ventilator frequency reduced to 50% and tidal volume reduced to 50% of normoventilation) at F10(2) 0.50 and at F10(2) 0.21. Hemodynamics, mixed venous O2-saturation, arterial and venous blood gases, and gas-exchanger exhaust were analyzed after stabilization over 15 minutes (mean +/- standard deviation). Blood gas analyses showed significant improvement with intravascular gas exchange during hypoventilation at F10(2) 0.21: pH moved from 7.10 +/- 0.17 to 7.19 +/- 0.15*, PaCO2 moved from 9.9 +/- 4.2 kPa to 8.7 +/- 2.8*, PaO2 moved from 6.5 +/- 1.2 kPa to 7.3 +/- 0.8* and mixed venous O2-saturation moved from 33.9 +/- 16.0% to 48.1 +/- 4.6* (* = p less than 0.05 for without versus with intravascular gas exchange).(ABSTRACT TRUNCATED AT 250 WORDS)
Mots-clé
Animals
Carbon Dioxide/*blood
Cattle
Oxygen/*blood
*Oxygenators, Membrane
Respiratory Insufficiency/blood/*therapy
Time Factors
Pubmed
Web of science
Création de la notice
14/02/2008 14:17
Dernière modification de la notice