Utility of the Platelet Function Analyzer in Patients with Suspected Platelet Function Disorders: Diagnostic Accuracy Study.

Détails

Ressource 1Télécharger: Thiemes-0040-1721502.pdf (334.21 [Ko])
Etat: Public
Version: Final published version
Licence: CC BY 4.0
ID Serval
serval:BIB_03605D5075AF
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Utility of the Platelet Function Analyzer in Patients with Suspected Platelet Function Disorders: Diagnostic Accuracy Study.
Périodique
TH open
Auteur⸱e⸱s
Kaufmann J., Adler M., Alberio L., Nagler M.
ISSN
2512-9465 (Electronic)
ISSN-L
2512-9465
Statut éditorial
Publié
Date de publication
10/2020
Peer-reviewed
Oui
Volume
4
Numéro
4
Pages
e427-e436
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: epublish
Résumé
Introduction The platelet function analyzer (PFA) is widely used as a screening tool for bleeding disorders in various settings. The diagnostic performance regarding platelet function disorders (PFDs), which are among the most common inherited bleeding disorders, is however still elusive. We aimed to assess the diagnostic value of PFA for PFD in clinical practice. Methods Comprehensive clinical and laboratory data of all consecutive patients referred to a specialized outpatient between January 2012 and March 2017 with a suspected bleeding disorder were prospectively recorded. The diagnostic work-up was performed according to a prespecified protocol following current guidelines and platelet function was tested using light transmission aggregometry as well as flow cytometry. Results Five hundred and fifty-five patients were included (median age 43.7 years; interquartile range [IQR] 29.3, 61.7; 66.9% female). Possible PFD was diagnosed in 64 patients (11.5%) and confirmed PFD in 54 patients (9.7%). In patients with confirmed PFD, median closure times were 107 seconds (ADP or adenosine diphosphate; IQR 89, 130) and 169 seconds (EPI; IQR 121, 211). In patients without bleeding disorders, PFA closure times were 96 seconds (ADP; IQR 83, 109) and 137 seconds (EPI; IQR 116, 158). The sensitivity was 19.5% in case of PFA ADP (95%CI 12.6, 30.0; specificity 86.4%; 95% CI 82.4, 89.8), and 44.3% in case of PFA EPI (95% CI 34.9, 53.9; specificity 75.6%; 95% CI 70.8, 79.9). Conclusion The diagnostic performance of PFA for PFD was moderate to poor. Our results do not support the utilization of PFA for screening of PFD in clinical practice.