Albuterol delivery in an in vitro pediatric ventilator lung model: comparison of jet, ultrasonic, and mesh nebulizers.

Détails

ID Serval
serval:BIB_02C34D2A1679
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Albuterol delivery in an in vitro pediatric ventilator lung model: comparison of jet, ultrasonic, and mesh nebulizers.
Périodique
Pediatric Critical Care Medicine
Auteur⸱e⸱s
Sidler-Moix A.L., Dolci U., Berger-Gryllaki M., Pannatier A., Cotting J., Di Paolo E.R.
ISSN
1529-7535 (Print)
ISSN-L
1529-7535
Statut éditorial
Publié
Date de publication
2013
Volume
14
Numéro
2
Pages
e98
Langue
anglais
Notes
Publication types: Journal ArticlePublication Status: ppublish
Résumé
OBJECTIVE: : To determine the influence of nebulizer types and nebulization modes on bronchodilator delivery in a mechanically ventilated pediatric lung model.
DESIGN: : In vitro, laboratory study.
SETTING: : Research laboratory of a university hospital.
INTERVENTIONS: : Using albuterol as a marker, three nebulizer types (jet nebulizer, ultrasonic nebulizer, and vibrating-mesh nebulizer) were tested in three nebulization modes in a nonhumidified bench model mimicking the ventilatory pattern of a 10-kg infant. The amounts of albuterol deposited on the inspiratory filters (inhaled drug) at the end of the endotracheal tube, on the expiratory filters, and remaining in the nebulizers or in the ventilator circuit were determined. Particle size distribution of the nebulizers was also measured.
MEASUREMENTS AND MAIN RESULTS: : The inhaled drug was 2.8% ± 0.5% for the jet nebulizer, 10.5% ± 2.3% for the ultrasonic nebulizer, and 5.4% ± 2.7% for the vibrating-mesh nebulizer in intermittent nebulization during the inspiratory phase (p < 0.01). The most efficient nebulizer was the vibrating-mesh nebulizer in continuous nebulization (13.3% ± 4.6%, p < 0.01). Depending on the nebulizers, a variable but important part of albuterol was observed as remaining in the nebulizers (jet and ultrasonic nebulizers), or being expired or lost in the ventilator circuit (all nebulizers). Only small particles (range 2.39-2.70 µm) reached the end of the endotracheal tube.
CONCLUSIONS: : Important differences between nebulizer types and nebulization modes were seen for albuterol deposition at the end of the endotracheal tube in an in vitro pediatric ventilator-lung model. New aerosol devices, such as ultrasonic and vibrating-mesh nebulizers, were more efficient than the jet nebulizer.
Pubmed
Web of science
Création de la notice
14/03/2013 19:06
Dernière modification de la notice
20/08/2019 13:24
Données d'usage