Octopus Standard Automated Perimetry, Pulsar Perimetry, Moorfields Motion Displacement Test and Heidelberg Retinal Tomography in Glaucoma Detection - A Comparison of Diagnostic Accuracy

Détails

ID Serval
serval:BIB_026472E02847
Type
Actes de conférence (partie): contribution originale à la littérature scientifique, publiée à l'occasion de conférences scientifiques, dans un ouvrage de compte-rendu (proceedings), ou dans l'édition spéciale d'un journal reconnu (conference proceedings).
Sous-type
Abstract (résumé de présentation): article court qui reprend les éléments essentiels présentés à l'occasion d'une conférence scientifique dans un poster ou lors d'une intervention orale.
Collection
Publications
Institution
Titre
Octopus Standard Automated Perimetry, Pulsar Perimetry, Moorfields Motion Displacement Test and Heidelberg Retinal Tomography in Glaucoma Detection - A Comparison of Diagnostic Accuracy
Titre de la conférence
Investigative Ophthalmology and Visual Science
Auteur⸱e⸱s
Oleszczuk J.D., Bergin C., Schnyder C., Crabb D., Sharkawi E.
Organisation
ARVO E-Abstract 4911/A334
Statut éditorial
Publié
Date de publication
2010
Peer-reviewed
Oui
Volume
51
Langue
anglais
Résumé
Purpose:
To investigate the accuracy of 4 clinical instruments in the detection of glaucomatous damage.
Methods:
102 eyes of 55 test subjects (Age mean = 66.5yrs, range = [39; 89]) underwent Heidelberg Retinal Tomography (HRTIII), (disc area<2.43); and standard automated perimetry (SAP) using Octopus (Dynamic); Pulsar (TOP); and Moorfields Motion Displacement Test (MDT) (ESTA strategy). Eyes were separated into three groups 1) Healthy (H): IOP<21mmHg and healthy discs (clinical examination), 39 subjects, 78 eyes; 2) Glaucoma suspect (GS): Suspicious discs (clinical examination), 12 subjects, 15 eyes; 3) Glaucoma (G): progressive structural or functional loss, 14 subjects, 20 eyes. Clinical diagnostic precision was examined using the cut-off associated with the p<5% normative limit of MD (Octopus/Pulsar), PTD (MDT) and MRA (HRT) analysis. The sensitivity, specificity and accuracy were calculated for each instrument.
Results:
See table
Conclusions:
Despite the advantage of defining glaucoma suspects using clinical optic disc examination, the HRT did not yield significantly higher accuracy than functional measures. HRT, MDT and Octopus SAP yielded higher accuracy than Pulsar perimetry, although results did not reach statistical significance. Further studies are required to investigate the structure-function correlations between these instruments.