Stent shortening and elongation: pitfalls with the Wiktor coronary endoprosthesis
Details
Serval ID
serval:BIB_FBC297999F58
Type
Article: article from journal or magazin.
Publication sub-type
Case report (case report): feedback on an observation with a short commentary.
Collection
Publications
Institution
Title
Stent shortening and elongation: pitfalls with the Wiktor coronary endoprosthesis
Journal
Catheterization and Cardiovascular Diagnosis
ISSN
1522-1946
0098-6569 (Print)
0098-6569 (Print)
Publication state
Published
Issued date
03/1994
Volume
31
Number
3
Pages
233-5
Notes
Case Reports
Journal Article --- Old month value: Mar
Journal Article --- Old month value: Mar
Abstract
This report demonstrates two specific complications that occurred during implantation of a Wiktor (Medtronic, Inc., Minneapolis, MN) coronary stent: stent shortening and elongation were observed and attributed to the unique coil structure of the device. Major cardiac events occurred during follow-up as a consequence of this stent deformation. Therefore, semiurgent bypass surgery should be considered to treat this complication.
Keywords
Aged
Angioplasty, Transluminal, Percutaneous Coronary/instrumentation
Coronary Angiography
Coronary Disease/*therapy
Equipment Design
Female
Humans
Male
Middle Aged
Myocardial Infarction/*etiology
Stents/*adverse effects
Thrombosis/*etiology
Pubmed