Results of the Patient-Related Outcomes of Mechanical lead Extraction Techniques (PROMET) study: a multicentre retrospective study on advanced mechanical lead extraction techniques.
Details
Serval ID
serval:BIB_EEA3C1640F17
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Results of the Patient-Related Outcomes of Mechanical lead Extraction Techniques (PROMET) study: a multicentre retrospective study on advanced mechanical lead extraction techniques.
Journal
Europace
ISSN
1532-2092 (Electronic)
ISSN-L
1099-5129
Publication state
Published
Issued date
01/07/2020
Peer-reviewed
Oui
Volume
22
Number
7
Pages
1103-1110
Language
english
Notes
Publication types: Journal Article ; Multicenter Study ; Research Support, Non-U.S. Gov't
Publication Status: ppublish
Publication Status: ppublish
Abstract
Several large studies have documented the outcome of transvenous lead extraction (TLE), focusing on laser and mechanical methods. To date there has been no large series addressing the results obtained with rotational lead extraction tools. This retrospective multicentre study was designed to investigate the outcomes of mechanical and rotational techniques.
Data were collected on a total of 2205 patients (age 66.0 ± 15.7 years) with 3849 leads targeted for extraction in six European lead extraction centres. The commonest indication was infection (46%). The targeted leads included 2879 pacemaker leads (74.8%), 949 implantable cardioverter-defibrillator leads (24.6%), and 21 leads for which details were unknown; 46.6% of leads were passive fixation leads. The median lead dwell time was 74 months [interquartile range (IQR) 41-112]. Clinical success was obtained in 97.0% of procedures, and complete extraction was achieved for 96.5% of leads. Major complications occurred in 22/2205 procedures (1%), with a peri-operative or procedure-related mortality rate of 4/2205 (0.18%). Minor complications occurred in 3.1% of procedures. A total of 1552 leads (in 992 patients) with a median dwell time of 106 months (IQR 66-145) were extracted using the Evolution rotational TLE tool. In this subgroup, complete success was obtained for 95.2% of leads with a procedural mortality rate of 0.4%.
Patient outcomes in the PROMET study compare favourably with other large TLE trials, underlining the capability of rotational TLE tools and techniques to match laser methods in efficacy and surpass them in safety.
Data were collected on a total of 2205 patients (age 66.0 ± 15.7 years) with 3849 leads targeted for extraction in six European lead extraction centres. The commonest indication was infection (46%). The targeted leads included 2879 pacemaker leads (74.8%), 949 implantable cardioverter-defibrillator leads (24.6%), and 21 leads for which details were unknown; 46.6% of leads were passive fixation leads. The median lead dwell time was 74 months [interquartile range (IQR) 41-112]. Clinical success was obtained in 97.0% of procedures, and complete extraction was achieved for 96.5% of leads. Major complications occurred in 22/2205 procedures (1%), with a peri-operative or procedure-related mortality rate of 4/2205 (0.18%). Minor complications occurred in 3.1% of procedures. A total of 1552 leads (in 992 patients) with a median dwell time of 106 months (IQR 66-145) were extracted using the Evolution rotational TLE tool. In this subgroup, complete success was obtained for 95.2% of leads with a procedural mortality rate of 0.4%.
Patient outcomes in the PROMET study compare favourably with other large TLE trials, underlining the capability of rotational TLE tools and techniques to match laser methods in efficacy and surpass them in safety.
Keywords
Aged, Aged, 80 and over, Defibrillators, Implantable/adverse effects, Device Removal, Humans, Middle Aged, Pacemaker, Artificial, Retrospective Studies, Treatment Outcome, Complication, Evolution extraction sheath, Laser lead extraction, Rotational lead extraction, Technical success