Automatic, localized in vivo adjustment of all first- and second-order shim coils.

Details

Serval ID
serval:BIB_E390593489C1
Type
Article: article from journal or magazin.
Collection
Publications
Title
Automatic, localized in vivo adjustment of all first- and second-order shim coils.
Journal
Magnetic Resonance in Medicine
Author(s)
Gruetter R.
ISSN
0740-3194 (Print)
ISSN-L
0740-3194
Publication state
Published
Issued date
1993
Volume
29
Number
6
Pages
804-811
Language
english
Notes
Publication types: Journal ArticlePublication Status: ppublish
Abstract
The implementation of a "fast, automatic shimming technique by mapping along projections" (FASTMAP) on a 2.1-Tesla whole-body system is described. The method provides a localized adjustment of all first- (X, Y, Z) and second-order (Z2, ZX, ZY, X2-Y2, 2XY) coils in 2 min. The time savings are achieved by mapping the magnetic field along six projections rather than from whole-imaging data sets. An analysis of noise error propagation suggests that in 64-ml volumes the residual linewidths obtained with the method are negligible when the signal-to-rms-noise ratio is above 30. The initial application of the method to localized 13C, 31P, and 1H spectroscopy of the human brain resulted in linewidths of approximately 2 Hz for 13C (144-ml volume), approximately 2.5 Hz for 31P (36-ml volume) and approximately 4 Hz for 1H (36-ml volume) with symmetric and reproducible lineshapes.
Keywords
Brain/anatomy & histology, Humans, Image Processing, Computer-Assisted, Magnetic Resonance Spectroscopy/instrumentation, Magnetic Resonance Spectroscopy/methods, Time Factors
Pubmed
Web of science
Create date
04/08/2010 16:28
Last modification date
20/08/2019 17:07
Usage data