Comparison of different invasive hemodynamic methods for AV delay optimization in patients with cardiac resynchronization therapy: implications for clinical trial design and clinical practice.

Details

Ressource 1Download: BIB_DA37D890024E.P001.pdf (975.36 [Ko])
State: Public
Version: Final published version
Serval ID
serval:BIB_DA37D890024E
Type
Article: article from journal or magazin.
Collection
Publications
Title
Comparison of different invasive hemodynamic methods for AV delay optimization in patients with cardiac resynchronization therapy: implications for clinical trial design and clinical practice.
Journal
International Journal of Cardiology
Author(s)
Whinnett Z.I., Francis D.P., Denis A., Willson K., Pascale P., van Geldorp I., De Guillebon M., Ploux S., Ellenbogen K., Haïssaguerre M., Ritter P., Bordachar P.
ISSN
1874-1754 (Electronic)
ISSN-L
0167-5273
Publication state
Published
Issued date
2013
Volume
168
Number
3
Pages
2228-2237
Language
english
Notes
Publication types: Journal Article ; Research Support, Non-U.S. Gov't Publication Status: ppublish
Abstract
BACKGROUND: Reproducibility and hemodynamic efficacy of optimization of AV delay (AVD) of cardiac resynchronization therapy (CRT) using invasive LV dp/dtmax are unknown.
METHOD AND RESULTS: 25 patients underwent AV delay (AVD) optimisation twice, using continuous left ventricular (LV) dp/dtmax, systolic blood pressure (SBP) and pulse pressure (PP). We compared 4 protocols for comparing dp/dtmax between AV delays: We assessed for dp/dtmax, LVSBP and LVPP, test-retest reproducibility of the optimum. Optimization using immediate absolute dp/dtmax had poor reproducibility (SDD of replicate optima=41 ms; R(2)=0.45) as did delayed absolute (SDD 39 ms; R(2)=0.50). Multiple relative had better reproducibility: SDD 23 ms, R(2)=0.76, and (p<0.01 by F test). Compared with AAI pacing, the hemodynamic increment from CRT, with the nominal AV delay was LVSBP 2% and LVdp/dtmax 5%, while CRT with pre-determined optimal AVD gave 6% and 9% respectively.
CONCLUSIONS: Because of inevitable background fluctuations, optimization by absolute dp/dtmax has poor same-day reproducibility, unsuitable for clinical or research purposes. Reproducibility is improved by comparing to a reference AVD and making multiple consecutive measurements. More than 6 measurements would be required for even more precise optimization--and might be advisable for future study designs. With optimal AVD, instead of nominal, the hemodynamic increment of CRT is approximately doubled.