Nonenhanced arterial spin labeled carotid MR angiography using three-dimensional radial balanced steady-state free precession imaging.
Details
Serval ID
serval:BIB_C759F8E2C4F0
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Nonenhanced arterial spin labeled carotid MR angiography using three-dimensional radial balanced steady-state free precession imaging.
Journal
Journal of Magnetic Resonance Imaging
ISSN
1522-2586 (Electronic)
ISSN-L
1053-1807
Publication state
Published
Issued date
2015
Peer-reviewed
Oui
Volume
41
Number
4
Pages
1150-1156
Language
english
Notes
Publication types: Journal Article Publication Status: ppublish
Abstract
PURPOSE: To optimize and preliminarily evaluate a three-dimensional (3D) radial balanced steady-state free precession (bSSFP) arterial spin labeled (ASL) sequence for nonenhanced MR angiography (MRA) of the extracranial carotid arteries.
MATERIALS AND METHODS: The carotid arteries of 13 healthy subjects and 2 patients were imaged on a 1.5 Tesla MRI system using an undersampled 3D radial bSSFP sequence providing a scan time of ∼4 min and 1 mm(3) isotropic resolution. A hybridized scheme that combined pseudocontinuous and pulsed ASL was used to maximize arterial coverage. The impact of a post label delay period, the sequence repetition time, and radiofrequency (RF) energy configuration of pseudocontinuous labeling on the display of the carotid arteries was assessed with contrast-to-noise ratio (CNR) measurements. Faster, higher undersampled 2 and 1 min scans were tested.
RESULTS: Using hybridized ASL MRA and a 3D radial bSSFP trajectory, arterial CNR was maximized with a post label delay of 0.2 s, repetition times ≥ 2.5 s (P < 0.05), and by eliminating RF energy during the pseudocontinuous control phase (P < 0.001). With higher levels of undersampling, the carotid arteries were displayed in ≤ 2 min.
CONCLUSION: Nonenhanced MRA using hybridized ASL with a 3D radial bSSFP trajectory can display long lengths of the carotid arteries with 1 mm(3) isotropic resolution. J. Magn. Reson. Imaging 2015;41:1150-1156. © 2014 Wiley Periodicals, Inc.
MATERIALS AND METHODS: The carotid arteries of 13 healthy subjects and 2 patients were imaged on a 1.5 Tesla MRI system using an undersampled 3D radial bSSFP sequence providing a scan time of ∼4 min and 1 mm(3) isotropic resolution. A hybridized scheme that combined pseudocontinuous and pulsed ASL was used to maximize arterial coverage. The impact of a post label delay period, the sequence repetition time, and radiofrequency (RF) energy configuration of pseudocontinuous labeling on the display of the carotid arteries was assessed with contrast-to-noise ratio (CNR) measurements. Faster, higher undersampled 2 and 1 min scans were tested.
RESULTS: Using hybridized ASL MRA and a 3D radial bSSFP trajectory, arterial CNR was maximized with a post label delay of 0.2 s, repetition times ≥ 2.5 s (P < 0.05), and by eliminating RF energy during the pseudocontinuous control phase (P < 0.001). With higher levels of undersampling, the carotid arteries were displayed in ≤ 2 min.
CONCLUSION: Nonenhanced MRA using hybridized ASL with a 3D radial bSSFP trajectory can display long lengths of the carotid arteries with 1 mm(3) isotropic resolution. J. Magn. Reson. Imaging 2015;41:1150-1156. © 2014 Wiley Periodicals, Inc.