Assessment of early-stage optic nerve invasion in retinoblastoma using high-resolution 1.5 Tesla MRI with surface coils: a multicentre, prospective accuracy study with histopathological correlation.

Details

Serval ID
serval:BIB_BDAFC24495A3
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Assessment of early-stage optic nerve invasion in retinoblastoma using high-resolution 1.5 Tesla MRI with surface coils: a multicentre, prospective accuracy study with histopathological correlation.
Journal
European Radiology
Author(s)
Brisse H.J., de Graaf P., Galluzzi P., Cosker K., Maeder P., Göricke S., Rodjan F., de Jong M.C., Savignoni A., Aerts I., Desjardins L., Moll A.C., Hadjistilianou T., Toti P., van der Valk P., Castelijns J.A., Sastre-Garau X.
Working group(s)
European Retinoblastoma Imaging Collaboration (ERIC)
ISSN
1432-1084 (Electronic)
ISSN-L
0938-7994
Publication state
Published
Issued date
2015
Volume
25
Number
5
Pages
1443-1452
Language
english
Notes
Publication types: Journal Article
Publication Status: ppublish
Abstract
OBJECTIVES: To assess the accuracy of high-resolution (HR) magnetic resonance imaging (MRI) in diagnosing early-stage optic nerve (ON) invasion in a retinoblastoma cohort.
METHODS: This IRB-approved, prospective multicenter study included 95 patients (55 boys, 40 girls; mean age, 29 months). 1.5-T MRI was performed using surface coils before enucleation, including spin-echo unenhanced and contrast-enhanced (CE) T1-weighted sequences (slice thickness, 2 mm; pixel size <0.3 × 0.3 mm(2)). Images were read by five neuroradiologists blinded to histopathologic findings. ROC curves were constructed with AUC assessment using a bootstrap method.
RESULTS: Histopathology identified 41 eyes without ON invasion and 25 with prelaminar, 18 with intralaminar and 12 with postlaminar invasion. All but one were postoperatively classified as stage I by the International Retinoblastoma Staging System. The accuracy of CE-T1 sequences in identifying ON invasion was limited (AUC = 0.64; 95 % CI, 0.55 - 0.72) and not confirmed for postlaminar invasion diagnosis (AUC = 0.64; 95 % CI, 0.47 - 0.82); high specificities (range, 0.64 - 1) and negative predictive values (range, 0.81 - 0.97) were confirmed.
CONCLUSION: HR-MRI with surface coils is recommended to appropriately select retinoblastoma patients eligible for primary enucleation without the risk of IRSS stage II but cannot substitute for pathology in differentiating the first degrees of ON invasion.
KEY POINTS: • HR-MRI excludes advanced optic nerve invasion with high negative predictive value. • HR-MRI accurately selects patients eligible for primary enucleation. • Diagnosis of early stages of optic nerve invasion still relies on pathology. • Several physiological MR patterns may mimic optic nerve invasion.
Pubmed
Web of science
Create date
15/01/2015 12:53
Last modification date
20/08/2019 16:31
Usage data