Superior venous drainage in the "LifeBox": a portable extracorporeal oxygenator with a self-expanding venous cannula.
Details
Serval ID
serval:BIB_B9E5A7AD8ED7
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Superior venous drainage in the "LifeBox": a portable extracorporeal oxygenator with a self-expanding venous cannula.
Journal
Perfusion
ISSN
1477-111X (Electronic)
ISSN-L
0267-6591
Publication state
Published
Issued date
07/2010
Peer-reviewed
Oui
Volume
25
Number
4
Pages
211-215
Language
english
Notes
Publication types: Journal Article
Publication Status: ppublish
Publication Status: ppublish
Abstract
In an experimental setting, the performance of the LifeBox, a new portable extracorporeal membrane oxygenator (ECMO) system suitable for patient transport, is presented. Standard rectilinear percutaneous cannulae are normally employed for this purpose, but have limited flow and pressure delivery due to their rigid structure. Therefore, we aimed to determine the potential for flow increase by using self-expanding venous cannulae.
Veno-arterial bypass was established in three pigs (40.6+/-5.1 kg). The venous line of the cardiopulmonary bypass was established by cannulation of the external jugular vein. The arterial side of the circulation was secured by cannulation of the common carotid artery. Two different venous cannulae (SmartCanula 18/36F 430mm and Biomedicus 19F) were examined for their functional integrity when used in conjunction with the centrifugal pump (500-3000 RPM) of the LifeBox system.
At 1500, 2000, 2500, and 3000 RPM, the blood flow increased steadily for each cannula, but remained higher in the self-expanding cannula. That is, the 19F rectilinear cannula achieved a blood flow of 0.93+/-0.14, 1.47+/-0.37, 1.9+/-0.68, and 1.5+/-0.9 l/min, respectively, and the 18/36F self-expanding cannula achieved 1.1+/-0.1, 1.9+/-0.33, 2.8+/-0.39 and 3.66+/-0.52 l/min. However, when tested for venous line pressure, the standard venous cannula achieved -29+/-10.7mmHg while the self-expanding cannula achieved -13.6 +/-4.3mmHg at 1500 RMP. As the RPM increased from 2500 to 3000, the venous line pressure accounted for -141.9+/-20 and -98+/-7.3mmHg for the 19F rectilinear cannula and -30.6+/-6.4 and -45+/-11.6mmHg for the self-expanding cannula.
The self-expanding cannula exhibited superior venous drainage ability when compared to the performance of the standard rectilinear cannula with the use of the LifeBox. The flow rate achieved was approximately 40% greater than the standard drainage device, with a maximal pump flow recorded at 4.3l/min.
Veno-arterial bypass was established in three pigs (40.6+/-5.1 kg). The venous line of the cardiopulmonary bypass was established by cannulation of the external jugular vein. The arterial side of the circulation was secured by cannulation of the common carotid artery. Two different venous cannulae (SmartCanula 18/36F 430mm and Biomedicus 19F) were examined for their functional integrity when used in conjunction with the centrifugal pump (500-3000 RPM) of the LifeBox system.
At 1500, 2000, 2500, and 3000 RPM, the blood flow increased steadily for each cannula, but remained higher in the self-expanding cannula. That is, the 19F rectilinear cannula achieved a blood flow of 0.93+/-0.14, 1.47+/-0.37, 1.9+/-0.68, and 1.5+/-0.9 l/min, respectively, and the 18/36F self-expanding cannula achieved 1.1+/-0.1, 1.9+/-0.33, 2.8+/-0.39 and 3.66+/-0.52 l/min. However, when tested for venous line pressure, the standard venous cannula achieved -29+/-10.7mmHg while the self-expanding cannula achieved -13.6 +/-4.3mmHg at 1500 RMP. As the RPM increased from 2500 to 3000, the venous line pressure accounted for -141.9+/-20 and -98+/-7.3mmHg for the 19F rectilinear cannula and -30.6+/-6.4 and -45+/-11.6mmHg for the self-expanding cannula.
The self-expanding cannula exhibited superior venous drainage ability when compared to the performance of the standard rectilinear cannula with the use of the LifeBox. The flow rate achieved was approximately 40% greater than the standard drainage device, with a maximal pump flow recorded at 4.3l/min.
Keywords
Animals, Cardiopulmonary Bypass/instrumentation, Catheterization, Central Venous/instrumentation, Extracorporeal Membrane Oxygenation/instrumentation, Extracorporeal Membrane Oxygenation/methods, Heart-Lung Machine, Oxygenators, Membrane, Swine, Venous Pressure