Analysis of joint laxity after total ankle arthroplasty: cadaver study.

Details

Serval ID
serval:BIB_B17E68394461
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Analysis of joint laxity after total ankle arthroplasty: cadaver study.
Journal
Clinical Biomechanics
Author(s)
Watanabe K., Crevoisier X.M., Kitaoka H.B., Zhao K.D., Berglund L.J., Kaufman K.R., An K.N.
ISSN
1879-1271[electronic]
Publication state
Published
Issued date
2009
Volume
24
Number
8
Pages
655-660
Language
english
Abstract
BACKGROUND: Clinical results of total ankle arthroplasty with early designs were disappointing. Recently-developed ankle prostheses have good mid-term results; however, limited information is available regarding effects of total ankle arthroplasty on ankle laxity.
METHODS: Eight cadaveric lower extremities were tested with a custom device which enabled measurement of multi-axial forces, moments, and displacement during applied axial, shear, and rotational loading. Tests consisted of anterior-posterior and medial-lateral translation and internal-external rotation of the talus relative to the tibia during axial loads on the tibia simulating body weight (700 N) and an unloaded condition (5 N). Tests were performed in neutral, dorsiflexion, and plantarflexion. Laxity was determined for the intact ankle, and following insertion of an unconstrained total ankle implant, comparing load-displacement curve.
FINDINGS: Laxity after total ankle arthroplasty did not approximate the normal ankle in most conditions tested. Displacement was significantly greater for total ankle arthroplasty in both posterior and lateral translation, and internal rotation, with 5 N axial loading, and anterior-posterior, medial-lateral translation, and internal-external rotation for 700 N axial loading. For the 700 N axial load condition, in the neutral ankle position, total anterior-posterior translation averaged 0.4 mm (SD 0.2 mm), but 6.0 mm (SD 1.5 mm) after total ankle arthroplasty (P<0.01). This study demonstrated more laxity in the replaced ankle than normal ankle for both unloaded and 700 N axially loaded conditions.
INTERPRETATION: These data indicate the increased responsibility of the ligaments for ankle laxity after total ankle arthroplasty and suggest the importance of meticulous ligament reconstruction with total ankle arthroplasty operations.
Keywords
Aged, Aged, 80 and over, Animals, Ankle Joint/surgery, Cadaver, Equipment Failure Analysis, Female, Humans, Joint Instability/etiology, Joint Instability/physiopathology, Joint Prosthesis/adverse effects, Male, Middle Aged, Movement, Range of Motion, Articular, Torque, Treatment Outcome
Pubmed
Web of science
Create date
21/01/2010 15:34
Last modification date
20/08/2019 16:20
Usage data