Channelized Hotelling observer correlation with human observers for low-contrast detection in liver CT images.

Details

Serval ID
serval:BIB_AEF4C87C0D64
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Channelized Hotelling observer correlation with human observers for low-contrast detection in liver CT images.
Journal
Journal of medical imaging
Author(s)
Ba A., Abbey C.K., Racine D., Viry A., Verdun F.R., Schmidt S., Bochud F.O.
ISSN
2329-4302 (Print)
ISSN-L
2329-4302
Publication state
Published
Issued date
04/2019
Peer-reviewed
Oui
Volume
6
Number
2
Pages
025501
Language
english
Notes
Publication types: Journal Article
Publication Status: ppublish
Abstract
Task-based image quality procedures in CT that substitute a human observer with a model observer usually use single-slice images with uniform backgrounds from homogeneous phantoms. However, anatomical structures and inhomogeneities in organs generate noise that can affect the detection performance of human observers. The purpose of this work was to assess the impact of background type, uniform or liver, and the viewing modality, single- or multislice, on the detection performance of human and model observers. We collected abdominal CT scans from patients and homogeneous phantom scans in which we digitally inserted low-contrast signals that mimicked a liver lesion. We ran a rating experiment with the two background conditions with three signal sizes and three human observers presenting images in two reading modalities: single- and multislice. In addition, channelized Hotelling observers (CHO) for single- and multislice detection were implemented and evaluated according to their degree of correlation with the human observer performance. For human observers, there was a small but significant improvement in performance with multislice compared to the single-slice viewing mode. Our data did not reveal a significant difference between uniform and anatomical backgrounds. Model observers demonstrated a good correlation with human observers for both viewing modalities. Human observers have very similar performances in both multi- and single-slice viewing mode. It is therefore preferable to use single-slice CHO as this model is computationally more tractable than multislice CHO. However, using images from a homogeneous phantom can result in overestimating image quality as CHO performance tends to be higher in uniform than anatomical backgrounds, while human observers have similar detection performances.
Keywords
channelized Hotelling observers, computed tomography, image quality, liver, observer performance
Pubmed
Create date
14/06/2019 17:18
Last modification date
12/03/2022 7:29
Usage data