Abstract
Valved stents are new land for cardiac surgeons even though they are being used more frequently by interventional disciplines. This paper presents simple steps to build a patient-specific pulmonary valved stent and its delivery device. The design concept was tested by random participants at a med-tech meeting. The valved stent is constructed by linking an endoprosthetic graft with a valved-jugular-vein. The delivery device is made from a modified 5-ml syringe. Of 72 participants, 66 (92%) built and 60 participants implanted the device successfully into the targeted pulmonary position via a trans-infundibular access.
Keywords
Animals, Blood Vessel Prosthesis, Blood Vessel Prosthesis Implantation/instrumentation, Blood Vessel Prosthesis Implantation/methods, Heart Valve Prosthesis, Heart Valve Prosthesis Implantation/instrumentation, Heart Valve Prosthesis Implantation/methods, Prosthesis Design, Pulmonary Valve, Stents, Sus scrofa, Suture Techniques