Direct-access valve replacement a novel approach for off-pump valve implantation using valved stents

Details

Serval ID
serval:BIB_AA70E903F2A8
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Direct-access valve replacement a novel approach for off-pump valve implantation using valved stents
Journal
Journal of the American College of Cardiology
Author(s)
Huber  C. H., Cohn  L. H., von Segesser  L. K.
ISSN
0735-1097
Publication state
Published
Issued date
07/2005
Peer-reviewed
Oui
Volume
46
Number
2
Pages
366-70
Notes
Journal Article --- Old month value: Jul 19
Abstract
OBJECTIVES: This study validates the off-pump antegrade transventricular route for ultrasound-guided direct-access aortic valved stent implantation. BACKGROUND: Direct-access aortic valved stent implantation offers numerous advantages over the remote-access percutaneous approach and may one day provide an alternative to surgical aortic valve replacement. METHODS: Valved stents were implanted off-pump in 12 pigs (68.5.0 +/- 7.3 kg) via direct-access transapical approach using a left-sided mini-thoracotomy and continuous ultrasonic and fluoroscopic guidance. Acute valved stent function was studied with intravascular and intracardiac ultrasound. All valved stents were tested in vitro before insertion. Macroscopic analysis was performed at necropsy. RESULTS: In 8 of 12 pigs, valved stents were delivered to the target site over the native aortic valve leaflets without interference of coronary blood flow and with good acute valve function. Two valved stents were deployed and supra-annularly occluded the coronary orifice, leading to fatal outcome. Two valved stents dislodged into the left ventricle, one because of size mismatch and one that failed to unfold correctly. CONCLUSIONS: Twelve pigs underwent deployment of a valved stent in the aortic position. Six valves observed for an average 4.5-h period showed satisfactory postimplantation valve function.
Keywords
Animals *Aortic Valve *Heart Valve Prosthesis *Heart Valve Prosthesis Implantation/instrumentation/methods *Stents Swine Thoracotomy/methods
Pubmed
Web of science
Open Access
Yes
Create date
29/01/2008 15:44
Last modification date
20/08/2019 16:14
Usage data