Correction of through-plane deformation artifacts in stimulated echo acquisition mode cardiac imaging.

Details

Serval ID
serval:BIB_9F5845A1EE74
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Correction of through-plane deformation artifacts in stimulated echo acquisition mode cardiac imaging.
Journal
Magnetic Resonance In Medicine
Author(s)
Fahmy A.S., Pan L., Stuber M., Osman N.F.
ISSN
0740-3194[print], 0740-3194[linking]
Publication state
Published
Issued date
2006
Volume
55
Number
2
Pages
404-412
Language
english
Notes
Publication types: Journal Article ; Research Support, N.I.H., Extramural ; Research Support, Non-U.S. Gov't Publication Status: ppublish
Abstract
Attempts to use a stimulated echo acquisition mode (STEAM) in cardiac imaging are impeded by imaging artifacts that result in signal attenuation and nulling of the cardiac tissue. In this work, we present a method to reduce this artifact by acquiring two sets of stimulated echo images with two different demodulations. The resulting two images are combined to recover the signal loss and weighted to compensate for possible deformation-dependent intensity variation. Numerical simulations were used to validate the theory. Also, the proposed correction method was applied to in vivo imaging of normal volunteers (n = 6) and animal models with induced infarction (n = 3). The results show the ability of the method to recover the lost myocardial signal and generate artifact-free black-blood cardiac images.
Keywords
Animals, Artifacts, Computer Simulation, Contrast Media, Dogs, Gadolinium DTPA, Heart/anatomy & histology, Humans, Image Enhancement/methods, Magnetic Resonance Imaging/methods, Myocardial Infarction/pathology
Pubmed
Web of science
Create date
02/03/2010 17:04
Last modification date
20/08/2019 16:05
Usage data