An improved trap design for decoupling multinuclear RF coils.
Details
Serval ID
serval:BIB_8C47EE9A14BA
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
An improved trap design for decoupling multinuclear RF coils.
Journal
Magnetic Resonance in Medicine
ISSN
1522-2594 (Electronic)
ISSN-L
0740-3194
Publication state
Published
Issued date
2014
Peer-reviewed
Oui
Volume
72
Number
2
Pages
584-590
Language
english
Notes
Publication types: Journal ArticlePublication Status: ppublish
Abstract
PURPOSE: Multinuclear magnetic resonance spectroscopy and imaging require a radiofrequency probe capable of transmitting and receiving at the proton and non-proton frequencies. To minimize coupling between probe elements tuned to different frequencies, LC (inductor-capacitor) traps blocking current at the (1) H frequency can be inserted in non-proton elements. This work compares LC traps with LCC traps, a modified design incorporating an additional capacitor, enabling control of the trap reactance at the low frequency while maintaining (1) H blocking.
METHODS: Losses introduced by both types of trap were analysed using circuit models. Radiofrequency coils incorporating a series of LC and LCC traps were then built and evaluated at the bench. LCC trap performance was then confirmed using (1) H and (13) C measurements in a 7T human scanner.
RESULTS: LC and LCC traps both effectively block interaction between non-proton and proton coils at the proton frequency. LCC traps were found to introduce a sensitivity reduction of 5±2%, which was less than half of that caused by LC traps.
CONCLUSION: Sensitivity of non-proton coils is critical. The improved trap design, incorporating one extra capacitor, significantly reduces losses introduced by the trap in the non-proton coil. Magn Reson Med 72:584-590, 2014. © 2013 Wiley Periodicals, Inc.
METHODS: Losses introduced by both types of trap were analysed using circuit models. Radiofrequency coils incorporating a series of LC and LCC traps were then built and evaluated at the bench. LCC trap performance was then confirmed using (1) H and (13) C measurements in a 7T human scanner.
RESULTS: LC and LCC traps both effectively block interaction between non-proton and proton coils at the proton frequency. LCC traps were found to introduce a sensitivity reduction of 5±2%, which was less than half of that caused by LC traps.
CONCLUSION: Sensitivity of non-proton coils is critical. The improved trap design, incorporating one extra capacitor, significantly reduces losses introduced by the trap in the non-proton coil. Magn Reson Med 72:584-590, 2014. © 2013 Wiley Periodicals, Inc.
Keywords
X-nucleus, RF coils, 13C, trap, dual-resonant