Differences in PO2 and PCO2 between arterial and arterialized earlobe samples

Details

Serval ID
serval:BIB_82B320C34C9A
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Differences in PO2 and PCO2 between arterial and arterialized earlobe samples
Journal
European Respiratory Journal
Author(s)
Sauty  A., Uldry  C., Debetaz  L. F., Leuenberger  P., Fitting  J. W.
ISSN
0903-1936 (Print)
Publication state
Published
Issued date
02/1996
Volume
9
Number
2
Pages
186-9
Notes
Comparative Study
Journal Article --- Old month value: Feb
Abstract
Arterialized ear lobe blood samples have been described as adequate to gauge gas exchange in acute and chronically ill patients. It is a safe procedure, usually performed by medical technicians. We have conducted a prospective study to verify the validity of this method. One hundred and fifteen consecutive adult patients were studied. Blood samples were drawn simultaneously from arterialized earlobe and radial artery. Values of partial pressure of oxygen (PO2) and of carbon dioxide (PCO2) were measured by means of blood gas electrodes. The correlation coefficients between the two samples were 0.928 for PO2 and 0.957 for PCO2 values. In spite of a highly significant correlation, the limits of agreement between the two methods were wide for PO2. Earlobe values of PO2 were usually lower than arterial values, with larger differences in the range of normal arterial PO2. On the other hand, the error and the limits of agreement were smaller for PCO2. We conclude that, in adult patients, arterialized earlobe blood PO2 is not a reliable mirror of arterial PO2.
Keywords
Adult Blood Gas Analysis/*methods Capillaries Carbon Dioxide/*blood Ear, External/*blood supply Humans Oxygen/*blood Prospective Studies Pulmonary Gas Exchange Radial Artery/physiology Reproducibility of Results