Fracture of a Transcatheter Atrial Septal Defect Occluder Device Causing Mitral Valve Perforation.

Details

Serval ID
serval:BIB_7E8931ADD0DC
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Fracture of a Transcatheter Atrial Septal Defect Occluder Device Causing Mitral Valve Perforation.
Journal
The Annals of thoracic surgery
Author(s)
Werner R.S., Prêtre R., Maisano F., Wilhelm M.J.
ISSN
1552-6259 (Electronic)
ISSN-L
0003-4975
Publication state
Published
Issued date
07/2019
Peer-reviewed
Oui
Volume
108
Number
1
Pages
e29-e30
Language
english
Notes
Publication types: Case Reports ; Journal Article ; Research Support, Non-U.S. Gov't
Publication Status: ppublish
Abstract
Transcatheter atrial septal defect (ASD) device closure has gained increasing popularity over the past decades due to shorter hospital stay and the absence of skin scars. However, concern about the seriousness of device-related complications is accumulating. We report a case of device fracture in a young asymptomatic woman almost 4 years after percutaneous secundum ASD closure, resulting in mitral valve perforation. Subsequently, elective surgical removal of the device and mitral valve reconstruction was performed. This case demonstrates that complications from transcatheter ASD closure may even occur late after implantation.
Keywords
Adult, Female, Heart Septal Defects, Atrial/surgery, Humans, Mitral Valve/injuries, Mitral Valve/surgery, Septal Occluder Device/adverse effects