Periprocedural safety and technical outcomes of the new Silk Vista Baby flow diverter for the treatment of intracranial aneurysms: results from a multicenter experience.

Details

Serval ID
serval:BIB_7D570B72AA45
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Periprocedural safety and technical outcomes of the new Silk Vista Baby flow diverter for the treatment of intracranial aneurysms: results from a multicenter experience.
Journal
Journal of neurointerventional surgery
Author(s)
Martínez-Galdámez M., Biondi A., Kalousek V., Pereira V.M., Ianucci G., Kaesmacher J., Gentric J.C., Tsang COA, Mosimann P.J., Brisbois D., Čulo B., Schob S., Quäschling U., Ognard J., Escartín J., Chabert E., Turjman F., Barbier C., Mihalea C., Spelle L., Chapot R.
ISSN
1759-8486 (Electronic)
ISSN-L
1759-8478
Publication state
Published
Issued date
07/2019
Peer-reviewed
Oui
Volume
11
Number
7
Pages
723-727
Language
english
Notes
Publication types: Journal Article ; Multicenter Study
Publication Status: ppublish
Abstract
The aim of our study was to assess the technical success and the safety of this new low-profile flow diverter Silk Vista Baby (SVB) by evaluating the intraprocedural and periprocedural complication rate.
Clinical, procedural, and angiographic data were analyzed.
41 consecutive patients (28 women; age average 50.5 years) with 43 aneurysms were treated with SVB. Aneurysm sizes were classified by their maximum diameter, with an average size of 9.5 mm (range 2-30 mm). Thirty-four cases were unruptured. five aneurysms previously ruptured, had recurrence after the initial coiling. There were two ruptured cases. Aneurysms' locations were: M1 segment (five cases), M2 segment (three cases), M3 segment (one case), middle cerebral artery (MCA) bifurcation (six cases), carotid-T (two cases), anterior communicating artery/A1/A2 (11 cases), pericallosal artery (four cases), supraclinoid ICA (two cases), PCom (one case), V4 segment (three cases), PCA (three cases), SCA (one case), and PICA (one case). We had five intraprocedural complications which resolved without clinical consequences and three events postprocedural events. Initial occlusion rates were: eight aneurysms (18.6%) were completely occluded, five aneurysms (11.6%) showed near-complete occlusion, four cases (9.3%) showed incomplete filling, and 26 cases (60.4%) showed persisting filling. The mRS score at discharge from the hospital did not change from the admission mRS score.
Our study demonstrated that the use of the new low-profile flow diverter, SVB device, for the treatment of intracranial aneurysms is feasible and technically safe.
Keywords
Adult, Aged, Angiography/methods, Angiography/trends, Anterior Cerebral Artery/diagnostic imaging, Anterior Cerebral Artery/surgery, Embolization, Therapeutic/methods, Endovascular Procedures/methods, Female, Humans, Intracranial Aneurysm/diagnostic imaging, Intracranial Aneurysm/surgery, Male, Middle Aged, Middle Cerebral Artery/diagnostic imaging, Middle Cerebral Artery/surgery, Perioperative Care/methods, Perioperative Care/trends, Retrospective Studies, Self Expandable Metallic Stents/adverse effects, Self Expandable Metallic Stents/trends, Treatment Outcome, aneurysm, flow diverter