Management of patient dose and image noise in routine pediatric CT abdominal examinations.

Details

Ressource 1Download: REF.pdf (224.28 [Ko])
State: Public
Version: Final published version
License: Not specified
It was possible to publish this article open access thanks to a Swiss National Licence with the publisher.
Serval ID
serval:BIB_766AE2E82D96
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Management of patient dose and image noise in routine pediatric CT abdominal examinations.
Journal
European Radiology
Author(s)
Verdun F.R., Lepori D., Monnin P., Valley J.F., Schnyder P., Gudinchet F.
ISSN
0938-7994
Publication state
Published
Issued date
2004
Peer-reviewed
Oui
Volume
14
Number
5
Pages
835-841
Language
english
Notes
Publication types: Evaluation Studies ; Journal Article - Publication Status: ppublish
Abstract
The aim was to propose a strategy for finding reasonable compromises between image noise and dose as a function of patient weight. Weighted CT dose index (CTDI(w)) was measured on a multidetector-row CT unit using CTDI test objects of 16, 24 and 32 cm in diameter at 80, 100, 120 and 140 kV. These test objects were then scanned in helical mode using a wide range of tube currents and voltages with a reconstructed slice thickness of 5 mm. For each set of acquisition parameter image noise was measured and the Rose model observer was used to test two strategies for proposing a reasonable compromise between dose and low-contrast detection performance: (1) the use of a unique noise level for all test object diameters, and (2) the use of a unique dose efficacy level defined as the noise reduction per unit dose. Published data were used to define four weight classes and an acquisition protocol was proposed for each class. The protocols have been applied in clinical routine for more than one year. CTDI(vol) values of 6.7, 9.4, 15.9 and 24.5 mGy were proposed for the following weight classes: 2.5-5, 5-15, 15-30 and 30-50 kg with image noise levels in the range of 10-15 HU. The proposed method allows patient dose and image noise to be controlled in such a way that dose reduction does not impair the detection of low-contrast lesions. The proposed values correspond to high- quality images and can be reduced if only high-contrast organs are assessed.
Keywords
Adolescent, Artifacts, Body Weight, Child, Child, Preschool, Clinical Protocols, Humans, Image Processing, Computer-Assisted, Infant, Infant, Newborn, Models, Theoretical, Radiation Dosage, Radiographic Image Enhancement, Radiography, Abdominal, Radiometry, Sensitivity and Specificity, Signal Processing, Computer-Assisted, Technology, Radiologic, Tomography, X-Ray Computed