Non-invasive monitoring of pulmonary artery pressure from timing information by EIT: experimental evaluation during induced hypoxia.
Details
Serval ID
serval:BIB_664A6D2B1CA6
Type
Article: article from journal or magazin.
Publication sub-type
Case report (case report): feedback on an observation with a short commentary.
Collection
Publications
Institution
Title
Non-invasive monitoring of pulmonary artery pressure from timing information by EIT: experimental evaluation during induced hypoxia.
Journal
Physiological Measurement
ISSN
1361-6579 (Electronic)
ISSN-L
0967-3334
Publication state
Published
Issued date
2016
Peer-reviewed
Oui
Volume
37
Number
6
Pages
713-726
Language
english
Abstract
Monitoring of pulmonary artery pressure (PAP) in pulmonary hypertensive patients is currently limited to invasive solutions. We investigate a novel non-invasive approach for continuous monitoring of PAP, based on electrical impedance tomography (EIT), a safe, low-cost and non-invasive imaging technology. EIT recordings were performed in three healthy subjects undergoing hypoxia-induced PAP variations. The pulmonary pulse arrival time (PAT), a timing parameter physiologically linked to the PAP, was automatically calculated from the EIT signals. Values were compared to systolic PAP values from Doppler echocardiography, and yielded strong correlation scores ([Formula: see text]) for all three subjects. Results suggest the feasibility of non-invasive, unsupervised monitoring of PAP.
Pubmed
Web of science
Create date
26/06/2016 14:41
Last modification date
20/08/2019 14:22