Accuracy of the smartphone blood pressure measurement solution OptiBP to track blood pressure changes in pregnant women.
Details
Serval ID
serval:BIB_65752B380A7C
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Accuracy of the smartphone blood pressure measurement solution OptiBP to track blood pressure changes in pregnant women.
Journal
Journal of hypertension
ISSN
1473-5598 (Electronic)
ISSN-L
0263-6352
Publication state
Published
Issued date
01/04/2025
Peer-reviewed
Oui
Volume
43
Number
4
Pages
665-672
Language
english
Notes
Publication types: Journal Article
Publication Status: ppublish
Publication Status: ppublish
Abstract
Hypertensive disorders present significant morbidity and mortality during pregnancy. Although ambulatory blood pressure measurement remains the standard of care for normotensive women, self-monitoring at home is increasingly prevalent. The widespread use of smartphones worldwide has sparked interest in mobile applications that leverage the built-in hardware for blood pressure estimation, yet few trials have assessed their accuracy.
This prospective, longitudinal and monocentric study evaluated the accuracy of the OptiBP algorithm against standard oscillometric blood pressure measurements in a sample of pregnant women. Patients scheduled for elective caesarean sections were enrolled during the preoperative anesthesia consultations. Paired blood pressure measurements using OptiBP and the reference method were obtained at multiple time-points in late pregnancy and the postpartum period. Agreement between methods was assessed using the AAMI/ESH/ISO 81060-2:2018 standard thresholds of 5 ± 8 mmHg for mean ± standard deviation of the error (criterion 1) and patient-specific standard deviation of the mean error (criterion 2) and represented graphically by Bland-Altman scatterplots.
Forty-eight women were enrolled of which 32 completed the protocol, yielding 338 total valid measurement pairs. Mean and standard deviation of the error were -1.78 ± 7.94 and 1.19 ± 7.59, and the patient-specific standard deviation of the mean error was 4.68 and 4.52, for SBP and DBP, respectively.
Compared with blood pressure measurements taken with an oscillometric device, OptiBP's blood pressure estimates meet the AAMI/ESH/ISO 81060-2:2018 criteria.
This prospective, longitudinal and monocentric study evaluated the accuracy of the OptiBP algorithm against standard oscillometric blood pressure measurements in a sample of pregnant women. Patients scheduled for elective caesarean sections were enrolled during the preoperative anesthesia consultations. Paired blood pressure measurements using OptiBP and the reference method were obtained at multiple time-points in late pregnancy and the postpartum period. Agreement between methods was assessed using the AAMI/ESH/ISO 81060-2:2018 standard thresholds of 5 ± 8 mmHg for mean ± standard deviation of the error (criterion 1) and patient-specific standard deviation of the mean error (criterion 2) and represented graphically by Bland-Altman scatterplots.
Forty-eight women were enrolled of which 32 completed the protocol, yielding 338 total valid measurement pairs. Mean and standard deviation of the error were -1.78 ± 7.94 and 1.19 ± 7.59, and the patient-specific standard deviation of the mean error was 4.68 and 4.52, for SBP and DBP, respectively.
Compared with blood pressure measurements taken with an oscillometric device, OptiBP's blood pressure estimates meet the AAMI/ESH/ISO 81060-2:2018 criteria.