Reproducibility and within-day variability of body fat masurements using segmental bipolar bioelectrical impedance in women

Details

Serval ID
serval:BIB_45CF0AC9836D
Type
Inproceedings: an article in a conference proceedings.
Publication sub-type
Abstract (Abstract): shot summary in a article that contain essentials elements presented during a scientific conference, lecture or from a poster.
Collection
Publications
Institution
Title
Reproducibility and within-day variability of body fat masurements using segmental bipolar bioelectrical impedance in women
Title of the conference
16th European Congress on Obesity (ECO), Geneva, Switzerland, 14-17 May 2008
Author(s)
Marques-Vidal Pedro Manuel, Paccaud Fred
ISBN
0307-0565
Publication state
Published
Issued date
2008
Volume
32
Series
International Journal of Obesity
Pages
S220
Language
english
Abstract
Objective: to assess the between and within-device reproducibility, as well as within-day variability of body fat measurements.
Methods: body fat percentage (%BF) was measured twice on seventeen female students aged between 18 and 20 with a body mass index of 21.9 22.6 kg/m2 (mean SD) using seven bipolar bioelectrical impedance devices (BF-306) according to the manufacturer's recommendations. Each student was also measured each hour between 7:00 and 22:00. Statistical analysis was conducted using a general linear model for repeated measurements.
Results: the correlation between first and second measurements was very high (Pearson r between 0.985 and 1.000, p<0.001), as well as the correlation between devices (Pearson r between 0.986 and 0.999, all p<0.001). Repeated measurements analysis showed no differences were between devices (F test=0.83, p=0.59) or readings (first vs. second: F test=0.12, p=0.74). Conversely, significant differences were found between assessment periods throughout the day, measurements made in the morning being lower than those made in the afternoon. Assuming an overall daily average of 100 (based on all measurements), the values were 95.8 3.2 (mean SD) at 8:00 versus 101.3 3.0 at 20:00, corresponding to a mean change of 2.2 1.1 in %BF (F test for repeated values=6.58, p<0.001).
Conclusions: the between and within-device reproducibility for measuring body fat is high, enabling the use of multiple devices in a single study. Conversely, small but significant changes in body fat measurements occur during the day, urging body fat measurements to be performed at fixed times.