ORACLE: An analytical approach for T1, T2, proton density, and off-resonance mapping with phase-cycled balanced steady-state free precession.

Details

Serval ID
serval:BIB_2E9E81084BF6
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
ORACLE: An analytical approach for T1, T2, proton density, and off-resonance mapping with phase-cycled balanced steady-state free precession.
Journal
Magnetic resonance in medicine
Author(s)
Plähn NMJ, Safarkhanlo Y., Açikgöz B.C., Mackowiak ALC, Radojewski P., Bonanno G., Peper E.S., Heule R., Bastiaansen JAM
ISSN
1522-2594 (Electronic)
ISSN-L
0740-3194
Publication state
In Press
Peer-reviewed
Oui
Language
english
Notes
Publication types: Journal Article
Publication Status: aheadofprint
Abstract
To develop and validate a novel analytical approach simplifying , , proton density (PD), and off-resonance quantifications from phase-cycled balanced steady-state free precession (bSSFP) data. Additionally, to introduce a method to correct aliasing effects in undersampled bSSFP profiles.
Off-resonant-encoded analytical parameter quantification using complex linearized equations (ORACLE) provides analytical solutions for bSSFP profiles. which instantaneously quantify , , proton density (PD), and . An aliasing correction formalism was derived to allow undersampling of bSSFP profiles. ORACLE was used to quantify , , PD, / , and based on fully sampled ( ) bSSFP profiles from numerical simulations and 3T MRI experiments in phantom and 10 healthy subjects' brains. Obtained values were compared with reference scans in the same scan session. Aliasing correction was validated in subsampled ( ) bSSFP profiles in numerical simulations and human brains.
ORACLE quantifications agreed well with input values from simulations and phantom reference values (R <sup>2</sup> = 0.99). In human brains, and quantifications when compared with reference methods showed coefficients of variation below 2.9% and 3.9%, biases of 182 and 16.6 ms, and mean white-matter values of 642 and 51 ms using ORACLE. The quantification differed less than 3 Hz between both methods. PD and maps had comparable histograms. The maps effectively identified cerebrospinal fluid. Aliasing correction removed aliasing-related quantification errors in undersampled bSSFP profiles, significantly reducing scan time.
ORACLE enables simplified and rapid quantification of , , PD, and from phase-cycled bSSFP profiles, reducing acquisition time and eliminating biomarker maps' coregistration issues.
Keywords
aliasing correction, analytical solutions, brain, phase‐cycled bSSFP, quantitative MRI