Accelerated T2 Mapping of the Lumbar Intervertebral Disc: Highly Undersampled K-Space Data for Robust T2 Relaxation Time Measurement in Clinically Feasible Acquisition Times.

Details

Serval ID
serval:BIB_2BFE43FAFF56
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Accelerated T2 Mapping of the Lumbar Intervertebral Disc: Highly Undersampled K-Space Data for Robust T2 Relaxation Time Measurement in Clinically Feasible Acquisition Times.
Journal
Investigative radiology
Author(s)
Raudner M., Schreiner M., Hilbert T., Kober T., Weber M., Windhager R., Trattnig S., Juras V.
ISSN
1536-0210 (Electronic)
ISSN-L
0020-9996
Publication state
Published
Issued date
11/2020
Peer-reviewed
Oui
Volume
55
Number
11
Pages
695-701
Language
english
Notes
Publication types: Journal Article ; Research Support, Non-U.S. Gov't
Publication Status: ppublish
Abstract
T2 mapping of the intervertebral disc (IVD) can depict quantitative changes reflecting biochemical change due to loss of glycosaminoglycan content. Conventional T2 mapping is usually performed using a 2-dimensional multi-echo-spin echo sequence (2D-MESE) with long acquisition times that are generally not compatible with clinical routine. This study investigates the applicability of GRAPPATINI, a T2 mapping sequence combining undersampling, model-based reconstruction, and parallel imaging, to offer clinically feasible acquisition times in T2 mapping of the lumbar IVD.
Fifty-eight individuals (26 female; mean age, 23.3 ± 8.1 years) were prospectively studied at 3 T. GRAPPATINI was conducted with the same parameters as the 2D-MESE while shortening the acquisition time from 13:18 to 2:27 minutes. The setup was also validated in a phantom experiment using a 6.48-hour-long single echo-spin echo sequence as reference. The IVDs were manually segmented on 4 central slices.
The median nucleus pulposus showed a strong Pearson correlation coefficient between T2GRAPPATINI and T2MESE (rp = 0.919; P < 0.001). There was also a significant correlation for the ventral (rp = 0.241; P < 0.001) and posterior (rp = 0.418; P < 0.001) annular regions.In the single spin-echo phantom experiment, the most accurate T2 estimation was achieved using T2GRAPPATINI with a median absolute deviation of 15.3 milliseconds as compared with T2MESE with 26.5 milliseconds.
GRAPPATINI facilitates precise T2 mapping at 3 T in accordance with clinical standards and reference methods using the same parameters while shortening acquisition times from 13:18 to 2:27 minutes with the same parameters.
Keywords
Adolescent, Adult, Female, Humans, Image Processing, Computer-Assisted, Intervertebral Disc/diagnostic imaging, Lumbar Vertebrae/diagnostic imaging, Magnetic Resonance Imaging/methods, Male, Phantoms, Imaging, Time Factors, Young Adult
Pubmed
Web of science
Create date
24/07/2020 11:57
Last modification date
16/04/2024 7:11
Usage data