How Low Can We Go in Contrast-Enhanced CT Imaging of the Chest?: A Dose-Finding Cadaver Study Using the Model-based Iterative Image Reconstruction Approach.

Details

Serval ID
serval:BIB_29AFBDB62DE9
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
How Low Can We Go in Contrast-Enhanced CT Imaging of the Chest?: A Dose-Finding Cadaver Study Using the Model-based Iterative Image Reconstruction Approach.
Journal
Academic Radiology
Author(s)
Mueck F.G., Roesch S., Scherr M., Fischer F., Geyer L., Peschel O., Maxien D., Grabherr S., Wirth S.
ISSN
1878-4046 (Electronic)
ISSN-L
1076-6332
Publication state
Published
Issued date
2015
Volume
22
Number
3
Pages
345-356
Language
english
Notes
Publication types: Journal Article Publication Status: ppublish
Abstract
RATIONALE AND OBJECTIVES: Dose reduction may compromise patients because of a decrease of image quality. Therefore, the amount of dose savings in new dose-reduction techniques needs to be thoroughly assessed. To avoid repeated studies in one patient, chest computed tomography (CT) scans with different dose levels were performed in corpses comparing model-based iterative reconstruction (MBIR) as a tool to enhance image quality with current standard full-dose imaging.
MATERIALS AND METHODS: Twenty-five human cadavers were scanned (CT HD750) after contrast medium injection at different, decreasing dose levels D0-D5 and respectively reconstructed with MBIR. The data at full-dose level, D0, have been additionally reconstructed with standard adaptive statistical iterative reconstruction (ASIR), which represented the full-dose baseline reference (FDBR). Two radiologists independently compared image quality (IQ) in 3-mm multiplanar reformations for soft-tissue evaluation of D0-D5 to FDBR (-2, diagnostically inferior; -1, inferior; 0, equal; +1, superior; and +2, diagnostically superior). For statistical analysis, the intraclass correlation coefficient (ICC) and the Wilcoxon test were used.
RESULTS: Mean CT dose index values (mGy) were as follows: D0/FDBR = 10.1 ± 1.7, D1 = 6.2 ± 2.8, D2 = 5.7 ± 2.7, D3 = 3.5 ± 1.9, D4 = 1.8 ± 1.0, and D5 = 0.9 ± 0.5. Mean IQ ratings were as follows: D0 = +1.8 ± 0.2, D1 = +1.5 ± 0.3, D2 = +1.1 ± 0.3, D3 = +0.7 ± 0.5, D4 = +0.1 ± 0.5, and D5 = -1.2 ± 0.5. All values demonstrated a significant difference to baseline (P < .05), except mean IQ for D4 (P = .61). ICC was 0.91.
CONCLUSIONS: Compared to ASIR, MBIR allowed for a significant dose reduction of 82% without impairment of IQ. This resulted in a calculated mean effective dose below 1 mSv.