Optimization of spatial resolution for peripheral magnetic resonance angiography.

Details

Serval ID
serval:BIB_271D61BF0FE6
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Optimization of spatial resolution for peripheral magnetic resonance angiography.
Journal
Academic Radiology
Author(s)
Tang A., Cloutier G., Therasse E., Beaudoin G., Qanadli S.D., Giroux M.F., Boussion N., de Guise J.A., Oliva V.L., Soulez G.
ISSN
1076-6332
Publication state
Published
Issued date
2007
Peer-reviewed
Oui
Volume
14
Number
1
Pages
54-61
Language
english
Abstract
RATIONALE AND OBJECTIVES: To determine optimum spatial resolution when imaging peripheral arteries with magnetic resonance angiography (MRA). MATERIALS AND METHODS: Eight vessel diameters ranging from 1.0 to 8.0 mm were simulated in a vascular phantom. A total of 40 three-dimensional flash MRA sequences were acquired with incremental variations of fields of view, matrix size, and slice thickness. The accurately known eight diameters were combined pairwise to generate 22 "exact" degrees of stenosis ranging from 42% to 87%. Then, the diameters were measured in the MRA images by three independent observers and with quantitative angiography (QA) software and used to compute the degrees of stenosis corresponding to the 22 "exact" ones. The accuracy and reproducibility of vessel diameter measurements and stenosis calculations were assessed for vessel size ranging from 6 to 8 mm (iliac artery), 4 to 5 mm (femoro-popliteal arteries), and 1 to 3 mm (infrapopliteal arteries). Maximum pixel dimension and slice thickness to obtain a mean error in stenosis evaluation of less than 10% were determined by linear regression analysis. RESULTS: Mean errors on stenosis quantification were 8.8% +/- 6.3% for 6- to 8-mm vessels, 15.5% +/- 8.2% for 4- to 5-mm vessels, and 18.9% +/- 7.5% for 1- to 3-mm vessels. Mean errors on stenosis calculation were 12.3% +/- 8.2% for observers and 11.4% +/- 15.1% for QA software (P = .0342). To evaluate stenosis with a mean error of less than 10%, maximum pixel surface, the pixel size in the phase direction, and the slice thickness should be less than 1.56 mm2, 1.34 mm, 1.70 mm, respectively (voxel size 2.65 mm3) for 6- to 8-mm vessels; 1.31 mm2, 1.10 mm, 1.34 mm (voxel size 1.76 mm3), for 4- to 5-mm vessels; and 1.17 mm2, 0.90 mm, 0.9 mm (voxel size 1.05 mm3) for 1- to 3-mm vessels. CONCLUSION: Higher spatial resolution than currently used should be selected for imaging peripheral vessels.
Keywords
Constriction, Pathologic, Magnetic Resonance Angiography/methods, Peripheral Vascular Diseases/diagnosis, Phantoms, Imaging, Software
Pubmed
Web of science
Create date
11/04/2008 12:20
Last modification date
20/08/2019 13:06
Usage data