Effect of 0.35 T and 1.5 T magnetic fields on superficial dose in MR-guided radiotherapy of laryngeal cancer.
Details
Serval ID
serval:BIB_1D2FD326829B
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Effect of 0.35 T and 1.5 T magnetic fields on superficial dose in MR-guided radiotherapy of laryngeal cancer.
Journal
Clinical and translational radiation oncology
ISSN
2405-6308 (Electronic)
ISSN-L
2405-6308
Publication state
Published
Issued date
05/2023
Peer-reviewed
Oui
Volume
40
Pages
100624
Language
english
Notes
Publication types: Journal Article
Publication Status: epublish
Publication Status: epublish
Abstract
Treatment of head and neck cancer on linear accelerators with on-board magnetic resonance imaging (MR-linac) might be beneficial to reduce side effects and increase accuracy. For many head and neck cancer patients, dose coverage of the often superficially located planning target volumes (PTVs) is required. This study examines the impact of the electron return effect (ERE) on the surface dose in MR-guided radiotherapy (MRgRT) compared to conventional radiotherapy.
For this bicentric dosimetric study, 14 cases of laryngeal carcinomas with PTVs reaching up to the skin surface were included. For each patient, five different plans were compared, two VMAT plans (with and without a 5 mm bolus) and three IMRT MRgRT plans (0.35 T, 1.5 T and 0 T, each without bolus). Dose distributions were also validated with film measurements.
A similar coverage on the most superficial 3-5 mm of the PTV was achieved in the VMAT plans with bolus and the MRgRT plans for both 0.35 T and 1.5 T. However, coverage on this region was usually not achieved for VMAT without bolus and the 0 T plans. The film measurements on phantoms confirmed the results with the relative error never exceeding the calculated differences between the plans.
The present study could demonstrate that the ERE for both commercially available MR-linac variants provides sufficient coverage of the superficial tissue layers in MRgRT-plans for laryngeal carcinoma.
For this bicentric dosimetric study, 14 cases of laryngeal carcinomas with PTVs reaching up to the skin surface were included. For each patient, five different plans were compared, two VMAT plans (with and without a 5 mm bolus) and three IMRT MRgRT plans (0.35 T, 1.5 T and 0 T, each without bolus). Dose distributions were also validated with film measurements.
A similar coverage on the most superficial 3-5 mm of the PTV was achieved in the VMAT plans with bolus and the MRgRT plans for both 0.35 T and 1.5 T. However, coverage on this region was usually not achieved for VMAT without bolus and the 0 T plans. The film measurements on phantoms confirmed the results with the relative error never exceeding the calculated differences between the plans.
The present study could demonstrate that the ERE for both commercially available MR-linac variants provides sufficient coverage of the superficial tissue layers in MRgRT-plans for laryngeal carcinoma.