Volumetric quantification of coronary artery calcifications using dual-slice spiral CT scanner: improved reproducibility of measurements with 180 degrees linear interpolation algorithm.

Details

Serval ID
serval:BIB_0ED673813224
Type
Article: article from journal or magazin.
Collection
Publications
Title
Volumetric quantification of coronary artery calcifications using dual-slice spiral CT scanner: improved reproducibility of measurements with 180 degrees linear interpolation algorithm.
Journal
Journal of Computer Assisted Tomography
Author(s)
Qanadli S.D., Mesurolle B., Aegerter P., Joseph T., Oliva V.L., Guertin M.C., Dubourg O., Fauchet M., Goeau-Brissonniére O.A., Lacombe P.
ISSN
0363-8715
Publication state
Published
Issued date
04/2001
Peer-reviewed
Oui
Volume
25
Number
2
Pages
278-86
Language
english
Abstract
PURPOSE: The purpose of this work was to determine the reproducibility of coronary total calcium score (TCS) with dual-slice helical CT and compare three acquisition protocols. METHOD: Fifty patients (59 +/- 10 years old) underwent dual-slice helical CT (collimation = 2 x 2.5 mm) and coronary angiography. Two successive scans were performed, resulting in three sets of images: pitch = 1, 360 degrees linear interpolation (LI) (A360); pitch = 1, 180 degrees LI (A180); and pitch = 1.5, 180 degrees LI (B180). TCS values, calculated using a volumetric method with a threshold of 90 HU, were compared, and the interscan variation was determined. Diagnostic performances were compared with receiver operating characteristic curves. RESULTS: Protocol A360 provided significantly lower TCS than protocols A180 and B180 (p < 0.0001). No statistical difference was seen between A180 and B180, which provided the lowest interscan variation (40 +/- 58%). However, no significant clinical impact of the observed interscan variations was found. CONCLUSION: Reproducibility of TCS with dual-slice helical CT is improved by the 180 LI algorithm. However, dual-slice helical CT is not sufficiently reproducible to allow serial quantification of TCS over time.
Keywords
Adult, Aged, Aged, 80 and over, Algorithms, Calcinosis/radiography, Cineangiography, Coronary Angiography, Coronary Artery Disease/radiography, Coronary Disease/radiography, Female, Humans, Male, Middle Aged, ROC Curve, Reproducibility of Results, Sensitivity and Specificity, Tomography Scanners, X-Ray Computed, Tomography, X-Ray Computed/methods
Pubmed
Web of science
Create date
09/04/2008 17:12
Last modification date
20/08/2019 13:35
Usage data